Biomechanical characterization of a novel ring connector for sutureless aortic anastomosis
Received:January 25, 2017  Revised:March 07, 2017
DOI10.7555/JBR.31.20170011
Keywordsaortic diseases, sutureless device, aortic anastomosis, swine
Grant Program
                       
AuthorInstitution
Huan Liu Department of Cardiovascular Surgery, The First Affiliated Hospital, Nanjing Medical University, Nanjing, Jiangsu , China
Shijiang Zhang Department of Cardiovascular Surgery, The First Affiliated Hospital, Nanjing Medical University, Nanjing, Jiangsu , China
Yongfeng Shao Department of Cardiovascular Surgery, The First Affiliated Hospital, Nanjing Medical University, Nanjing, Jiangsu , China
Xiaohu Lu Department of Cardiovascular Surgery, The First Affiliated Hospital, Nanjing Medical University, Nanjing, Jiangsu , China
Weidong Gu Department of Cardiovascular Surgery, The First Affiliated Hospital, Nanjing Medical University, Nanjing, Jiangsu , China
Buqing Ni Department of Cardiovascular Surgery, The First Affiliated Hospital, Nanjing Medical University, Nanjing, Jiangsu , China
Qun Gu Department of Cardiovascular Surgery, The First Affiliated Hospital, Nanjing Medical University, Nanjing, Jiangsu , China
Junjie Du Department of Cardiovascular Surgery, The First Affiliated Hospital, Nanjing Medical University, Nanjing, Jiangsu , China
Hits: 41
Download times: 32
      
Abstract
      The surgical treatment for aortic diseases remains a challenge for any cardiac surgeon. The use of sutureless ring connector in aortic anastomosis can simplify the procedure and shorten anastomosis time. Therefore, we developed a novel device for sutureless aortic anastomosis. A series of experiments were carried out for tensile and leakproofcapacity assessments to verify the feasibility of the ring connector by using fresh swine aorta samples. In in vivo test, the ring connector was implanted in 6 swine with follow-up of 6 months. Radiographic and pathological studies of the aorta were performed. In the tensile tests, the strength was 32.75.9 Newton (N) in the sutureless anastomosis group, compared with 73.312.5 N in the control group by traditional manual suture. In the leakproof-capacity assessment, no sign of either leakage or bursting was evident at 280 mmHg of internal pressure in the aorta samples. In in vivo tests, it took 9.470.3 minutes for the sutureless anastomosis, compared with 15.581.39 minutes for hand-sewn suturing. Insertion was easy and rapid. Radiographic and pathological studies were performed at first month, third month and sixth month after surgery, each time obtained from the two swine, showed patency of the anastomosis and no signs of stenosis, blood leakage, migration or pseudoaneurysm formation, except one paralyzed swine developed of thrombo-occlusion at the site of the sutureless anastomosis. The result indicates that this novel ring connector offers considerable promise for sutureless aortic anastomosis.